Sensor Type Setting And Calibration; Standard; Custom; Calibration - Unipulse TM201 Operation Manual

Torque monitor
Hide thumbs Also See for TM201:
Table of Contents

Advertisement

4 Calibration
4-2.

Sensor type setting and calibration

Standard

Select the type of UTMⅡ connected.
Chapter
4

Custom

Select UTMⅡ/UTM and input the rated capacity.
The rated capacity unit is Nm.
If using UTMⅡ not in the standard list, select custom.

Calibration

By setting the sensor type or clicking the "Calibration"
button, the following dialog is displayed. Click "OK" to
start calibration.
Once the calibration starts, display conditions of torque,
rotation speed and power can be selected in order.
Selection screens are displayed in the following order.
Select the appropriate condition respectively and click
"OK".
Click "Return" to go back to the previous selection screen and "Cancel" to stop the calibration.
Setting cannot be changed once canceled.
26
26
Key points
When the calibration setting protection is ON, calibration cannot be performed.
Follow (1) and (7) of P.29 "4-3.Procedures of equivalent input calibration and
actual load calibration (torque)" to change the setting.