4 Calibration
4-4.

Display setting of rotation speed/power

Settings related to display of the rotation speed and power are performed.
(1) Pulse rate (rotation speed) setting
Select the pulse input rate.
Select the number of pulses per rotation.
(2) Min. scale division (rotation speed) setting
Select the min. scale division (display resolution, scale
Chapter
division) of the rotation speed.
4
(3) Unit (power) setting
Select the unit of power.
The decimal place is not automatically changed.
(4) Decimal place (power) setting
Select the decimal place of the power.
The unit is not automatically changed.
(5) Min. scale division (power) setting
Select the min. scale division (display resolution, scale
division) of the power.
34
34
<Setting value>
1 pulse, 4 pulses
<Setting value>
1, 2, 5, 10
<Setting value>
mW, W, kW, PS, HP
<Setting value>
99999, 9999.9, 999.99, 99.999
<Setting value>
1, 2, 5, 10
Key points
The power is calculated according to the internal conversion based on the unit
and decimal place setting. Therefore, do not change these settings without due
consideration. Perform re-calibration to change these settings.
Refer to P.37 "4-6.Unit conversion table" for unit conversion table.