3 Setting Procedures
3-6.

Calibration tab

Setting of sensor type and setting and calibration related to the TM201 calibration values are
performed.
Chapter
3

Detail

When this item is checked, detail items related to calibration (calibration setting,
calibration, rotation speed) are displayed. If this check is removed, this item is
not displayed (default).

Sensor type

Select the type of UTMⅡ connected and start calibration.
Refer to P.26 "4-2.Sensor type setting and calibration" for
details.

Retrieve calibration settings

"Retrieve" button
Calibration setting values are received from the TM201 all at
once.
(Normally, these values are received at the time of this application startup.)

Initialization of calibration settings

"Initialize" button
Calibration setting values inside the TM201 are initialized.
When this button is clicked, a confirmation dialog is displayed. With this operation, settings on the
PC and inside theTM201 are initialized and cannot be restored; therefore, click the "Yes (Y)" button
only when it is absolutely necessary to initialize.
When the calibration setting protection is ON, initialization cannot be performed.
Refer to P.44 "7-1.List of setting values" for default values.
Calibration itself is not performed.