Inspection and maintenance
2.2.9
Arm system and single tank
Fig. 11:
SIM
Movement of the arm system and the single tank
•
The entire arm system including the single tank must not move by itself in any position.
However, it must be easy to move to a new position. If this is not the case, proceed ac-
cording to the section "Adjusting the friction linings" in the service instructions.
SIM
Movement of the single tank fork
•
Check that the single tank fork can be easily turned and stays in any desired position.
SIM
Single tank movement
•
Check that the single tank moves easily in the fork and stays in any desired position.
Siemens AG
Medical Solutions
SPR8-215.831.04.03.02
07.05
MOBILETT Plus/ -E/ -M
Page 31 of 64
CS PS 24
31