Inspection and maintenance
2.2.9

Arm system and single tank

Fig. 11:
SIM
Movement of the arm system and the single tank
The entire arm system including the single tank must not move by itself in any position.
However, it must be easy to move to a new position. If this is not the case, proceed ac-
cording to the section "Adjusting the friction linings" in the service instructions.
SIM
Movement of the single tank fork
Check that the single tank fork can be easily turned and stays in any desired position.
SIM
Single tank movement
Check that the single tank moves easily in the fork and stays in any desired position.
Siemens AG
Medical Solutions
SPR8-215.831.04.03.02
07.05
MOBILETT Plus/ -E/ -M
Page 31 of 64
CS PS 24
31