Inspection and maintenance
If the deviation is larger than 1.7%, correct the coincidence of the light and radiation
fields by adjusting the nuts marked "D" and "E".
X
Check the setting by releasing another exposure.
If necessary, repeat the adjustment until the deviation of the light and radiation fields
lies within the acceptable tolerance.
QSQ
Deviation ((A + C) / SID)
QSQ
Deviation ((B + D) / SID)