Inspection and maintenance
•
If the deviation is larger than 1.7%, correct the coincidence of the light and radiation
fields by adjusting the nuts marked "D" and "E".
•
X
Check the setting by releasing another exposure.
•
If necessary, repeat the adjustment until the deviation of the light and radiation fields
lies within the acceptable tolerance.
QSQ
Deviation ((A + C) / SID)
QSQ
Deviation ((B + D) / SID)
Siemens AG
Medical Solutions
SPR8-215.831.04.03.02
07.05
(Fig. 30 / p. 51)
MOBILETT Plus/ -E/ -M
Page 55 of 64
CS PS 24
55