Inspection and maintenance
2.20

Checking the kV and mAs displays

Check whether the green "ready" lamp lights up approximately 2 minutes after the
system is switched on. If it doesn't light up, an error message will appear on the dis-
play. Refer to the chapter "Display messages" in the service instructions.
Check whether the kV and mAs values are clearly visible from the operator's position.
Determine whether all exposure parameters are displayed correctly.
PMF
"Ready" indicator
SIE
Visibility of displays
SIE
Correct displays
Siemens AG
Medical Solutions
SPR8-215.831.04.03.02
07.05
MOBILETT Plus/ -E/ -M
Page 61 of 64
CS PS 24
61