Visual Inspection; Damage - Siemens MOBILETT Plus Maintenance Instructions Manual

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

24
2-
2
Inspection and maintenance
2.1

Visual inspection

SIM
Labels
Certification labels, labels with the model and serial numbers and a "WARNING" label
are attached to the unit. The label locations are indicated in the Instructions for Use.
Ensure that all labels are legible and visible.
PMP
Customer documentation
Check that the following documents are filed on site:
- Instructions for Use
- User statements and a copy of FDA form FD 2579, completed and signed by the as-
sembler at the time of installation
- Periodic maintenance records/checks/protocol
- Assembler reports after major repairs
2.1.1

Damage

Check the following parts for cracks and other damage.
Repair or replace any damaged parts.
SIM
Covers
SIM
Cassette compartment
SIM
Control and display panel
SIM
Exposure release cable
SIM
Single tank generator
SIM
Collimator
Make sure that neither the single tank generator nor the collimator have any mechani-
cal damage that could affect radiation protection.
SIM
DAP (optional)
Make sure there are no scratches on the DAP ionization chamber. Make sure the cable
and the connector are intact. Make sure the DAP display is intact as well.
MOBILETT Plus/ -E/ -M
SPR8-215.831.04.03.02
07.05
Inspection and maintenance
Page 24 of 64
CS PS 24
Siemens AG
Medical Solutions

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobilett plus eMobilett plus m

Table of Contents