Service Mode - Siemens MOBILETT Plus Maintenance Instructions Manual

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

38
2.4

Service mode

Fig. 20:
Reattach all cover parts with the exception of the upper system cover.
NOTE
Entering service mode
Switch S1:1 on board D1 must be switched over (S1:1 to ON) to call up the service pro-
gram. The program version is then displayed:
kV
vEr
(depends on the software version installed)
(To exit the service mode, switch S1:1 on board D1 must be switched back to OFF. This
will cause any program that is still running to be shut down.)
Selection of test programs with buttons "kV+/ -"
Several test programs are available. The respective program number appears in the kV
window of the display while the mAs window remains blank. Example of the display for
program no. 6:
kV
P06
MOBILETT Plus/ -E/ -M
Switch S10 on D7 OFF prior to the following inspection.
mAs
x.x = version x.x
mAs
SPR8-215.831.04.03.02
07.05
Inspection and maintenance
Page 38 of 64
CS PS 24
Siemens AG
Medical Solutions

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobilett plus eMobilett plus m

Table of Contents