32
2.2.10
Mains cable
Fig. 12:
Pos. 1
PMP
Replace the cable winder every 4 years (units with cable winder only).
•
The cable winder should be replaced every four years.
SIM
Cable winder (for units with cable winder only)
•
Move the cable up and down while winding it, in order to avoid any slack, until the cable
is completely rolled up. Ensure proper functioning of cable winder and brake.
PMP
Replace the mains cable every 2 years.
•
The mains cable must be replaced every two years.
NOTE
SIM
Damage
•
Check the mains cable and plug for damage. If necessary, replace them.
PMP
Cleaning
•
Clean the mains cable using a paper towel dampened with a lukewarm solution of wa-
ter and a household cleaning agent.
SIM
Protective conductor
•
Disconnect the power supply plug.
MOBILETT Plus/ -E/ -M
Protective conductor
New cable winder 65 08 746 comes with the new mains cable.
Therefore, it is not necessary to replace the mains cable, if the ca-
ble winder has been replaced already.