General
Technical Safety Checks,
Test certificate - check
6. Product-specific checks
7. Test result/evaluation:
Siemens AG
Medical Solutions
SPR8-215.831.04.03.02
07.05
Maintenance instructions - check/
Comments
2.2.7 Handles, check: SIM Mounting
2.2.8 Collimator adjustment knobs, check:
SIM Mounting
2.2.11 Arm system mounting, check:
SIM Screws and lock nuts
2.2.11.1 Adjustment screw, checks:
SIM Lock nuts
SIM Base mounting
2.2.11.2 Arm connecting piece, check:
SIM Screws and nuts
2.2.11.3 Single tank fork, checks:
SIM Screw connections on the fork
SIM Single tank screw connections
2.10 Remote exposure switch system
(optional), check:
SIE Function of remote exposure switch
system
2.11 DAP measuring system (optional),
check:
SIE Function of DAP measuring system
Evaluate the long-term trends of the pro-
tective conductor resistance and the equiv-
alent leakage current by comparing the
current measuring values with those of the
preceding technical safety check. A sud-
den or unexpected increase in the mea-
sured values may indicate a
safety-relevant defect – even if no limit
value is exceeded.
Fill in the result of the check.
Have the operator, or a person authorized
by the operator, sign the TSC protocol.
Hand the TSC protocol over to the opera-
tor, or a person authorized by the operator.