Inspection and maintenance
Pressing the "kV+" button calls up the next test program. As soon as the last test program
has been reached and "kV+" is pressed again, the display begins again with the program
version.
The "kV-" button allows you to run the test programs in reverse order. The "collimator
lamp" button is used for the ON/OFF function, i.e., it starts or stops the displayed test pro-
gram.
Example of the display for P06 after activation of the collimator lamp (starts the formation
of the capacitor bank):
kV
CAP
The test program is stopped by pressing the "collimator lamp" button a second time.
Example of the display for P06:
kV
P06
Program P06 is now stopped. A new service program can be selected with the "kV+/-"
buttons or program no. 6 can be restarted by pressing the "collimator lamp" button.
NOTE
Siemens AG
Medical Solutions
mAs
For = capacitor formation program
mAs
Stop
Messages appearing in service mode, with the exception of "Quick
adjustment", are only displayed, i.e. they are not incorporated in
the error memory.
SPR8-215.831.04.03.02
07.05
MOBILETT Plus/ -E/ -M