Inspection and maintenance
2.12

Replacing the collimator lamp

Fig. 29:
Fig. 31:
PMP
Replace every year
The collimator lamp must be replaced every year.
Remove the cover
(Fig. 29 / p.
Remove the protective plate
the screws marked "B" by one turn. Do not remove them. Do not remove nuts "C", "D"
and "E" in this process
Remove the lamp and replace it with a new lamp,
glass bulb with your fingers.
Siemens AG
Medical Solutions
(1/Fig. 29 / p. 51)
51). Turn the collimator so that the lamp housing is accessible.
(2/Fig. 30 / p. 51)
(Fig. 30 / p.
SPR8-215.831.04.03.02
07.05
Fig. 30:
by removing the four screws marked "A"
covering the collimator lamp. Loosen
51).
(Fig. 31 / p.
Page 51 of 64
CS PS 24
51). Do not touch the
MOBILETT Plus/ -E/ -M
51