Inspection and maintenance
2.12
Replacing the collimator lamp
Fig. 29:
Fig. 31:
PMP
Replace every year
•
The collimator lamp must be replaced every year.
•
Remove the cover
(Fig. 29 / p.
•
Remove the protective plate
the screws marked "B" by one turn. Do not remove them. Do not remove nuts "C", "D"
and "E" in this process
•
Remove the lamp and replace it with a new lamp,
glass bulb with your fingers.
Siemens AG
Medical Solutions
(1/Fig. 29 / p. 51)
51). Turn the collimator so that the lamp housing is accessible.
(2/Fig. 30 / p. 51)
(Fig. 30 / p.
SPR8-215.831.04.03.02
07.05
Fig. 30:
by removing the four screws marked "A"
covering the collimator lamp. Loosen
51).
(Fig. 31 / p.
Page 51 of 64
CS PS 24
51). Do not touch the
MOBILETT Plus/ -E/ -M
51