Inspection and maintenance
2.2.5

Brakes

Fig. 8:
WARNING
SIM
Uniformity
SIM
Braking force
SIM
Locking
Check whether the hand brake is sufficient to stop the unit and that it does not pull to the
left or the right when the brake is applied.
Lock the brake.
Remove the front, rear, left and right covers.
The brake force can be increased by increasing the tension on the brake cable
(Fig. 8 / p.
hand brake/parking brake" in the service instructions.
Siemens AG
Medical Solutions
There is a risk of life-threatening electric shock.
Make sure protective measures have been taken prior to
opening the unit.
Disconnect the system from the mains.
(Protective measures when working on the capacitor
bank / p.
It should not be possible to push the MOBILETT using average force. Slight
skidding is permissible when applying a force greater than 350 N.
29). Additional information can be found in the section on "Adjustment of the
SPR8-215.831.04.03.02
07.05
14).
Page 29 of 64
CS PS 24
29
MOBILETT Plus/ -E/ -M

Hide quick links: