During Recording
Paused Readings
During configuration of the UTREL30-16 you can set the option to ignore up to 15 readings
for alarm and statistics calculations after either button is pressed. The readings are still
shown on the graph and in the data listing, but they are Annabelle as paused, and their
value is ignored when determining alarm trigger conditions, minimum/maximum values
and other statistical calculations.
The Paused Readings feature is useful, for example, when you need to temporarily
remove the probe from the monitored location to inspect goods, but you do not
wish to trigger an alarm as a result of you handling the probe.
It allows the logger to acclimatize to the environment
again, before further readings are processed.
After a button press the display shows
the
symbol.
PAUSED
ignored reading has been recorded.
The option is set in the
configuration with LogTag
number of readings after the last button press.
Paused readings are specially marked in the graph and data
listings.
How long
PAUSED
button. It will show longer, if the button is pressed just after the logger takes a
reading, but shorter if you press the button just before. For example, if you
configure a logging interval of 10 minutes and 2 paused readings, the time
is shown could be as short as 10 minutes, but as long as 20 minutes.
Marking a reading with an inspection mark
When you press the
reading taken will be identified in the downloaded data and on the PDF report with an
inspection mark.
If the
Allow stopping with the Stop button
registered when you press the
stopping the logger. An inspection mark will also be recorded if you clear an alarm.
PAUSED
will turn off as soon as the last
Advanced Settings tab
®
Analyzer and is expressed in
is displayed depends on when between readings you press the
button while the UTREL30-16 is recording, the next
REVIEW
/MARK
START
/CLEAR/STOP
UTREL30-16 User Guide
next to
during
feature is enabled, a mark will also be
button, but do not complete the process of
36
PAUSED
Need help?
Do you have a question about the UTREL30-16 and is the answer not in the manual?
Questions and answers