Reviewing Day Statistics on the Display
Data from 7 days ago
After pressing the
minimum statistics for day -6) the maximum
statistic from seven days ago is displayed:
The
flashes to indicate that the data being
l
displayed is from seven days ago.
Seven days ago recordings above the upper
l
threshold were recorded, indicated by the upper
alarm threshold marker.
The duration was 5 hours 20 minutes.
l
An alarm event was generated, indicated by the
l
symbol. Neither duration from day -6 or day -
7 alone would cause an alarm event, however
the combined duration is 11 hours 50 minutes,
which is more than the allowed 10 accumulative
hours.
The maximum temperature recorded seven days
l
ago was -64.1 °C.