Audible Alarm
Audible Alarm
The UTREL30-16 is fitted with a buzzer. You can choose to activate the buzzer when an
alarm event has been triggered to provide extra feedback. This is enabled or disabled in the
Advanced Settings
when configuring the logger with LogTag
Please note, that continual activation of the audible alarm will reduce the working
life of the battery. When an alarm event is triggered, the alarm should be cleared
as soon as possible.
The alarm will sound once every four seconds for the first 24 hours, then sound less
frequently to preserve battery life
configured.
Note: The interval will increase, even if the alarm is re-triggered!
Please see more information about re-triggering alarms in
page 17
The buzzer will temporarily turn off when you are reviewing data and in Low Power
mode.
4
The interval at which the alarm sounds will increase to 8 seconds when the clock passes midnight for a second time (i.e. the
alarm will sound every 4 seconds between 24 and 48 hours, depending on when during the day the alarm was first triggered).
When the clock passes midnight a third time, the interval changes to 12 seconds until the alarm is cleared.