Clearing The Min/Max Values - LogTag Recorders UTREL30-16 Product User Manual

30-day low-temperature data logger with display
Table of Contents

Advertisement

During Recording

Clearing the Min/Max Values

The UTREL30-16 tracks the minimum and maximum temperatures since it started logging
and displays them below the last recorded temperature.
These values can be reset any time while the unit is
recording (but not once the unit has stopped).
To reset the values, follow this procedure:
 1. Press and hold the
 2. After 1 second the number values alternate with
dashes, after 4 seconds, the dashes remain on.
 3. Release the buttons within 2 seconds. The
values will be reset and the dashes remain on
screen.
The values will not be reset, and the currently stored min/max values will be retained if
 
you keep holding the button for more than 2 seconds after the dashes no longer
l
blink; or
 
you release the button within the initial 4 second period after the initial press.
l
In these cases, the display keeps showing the normal recording screen.
The logger will now track new min/max values. A min/max reset mark will be recorded in
the logged data. As soon as the next reading is taken, both minimum and maximum values
will be shown again.
You will be able to see on the PDF and in the summary in LogTag
min/max values were cleared, but you will not be able to review previous min/max values
on screen once they have been cleared.
UTREL30-16 User Guide