Standard Configuration Options

TREL30-16 Start Options

During configuration with LogTag
taking temperature readings:
Push button start
The logger will start taking temperature readings as soon as you have pressed the
button (see
START
/CLEAR/STOP
When you choose the push button start option, you can
disable it) and also
Pre-start readings
If you enable pre-start readings, the TREL30-16 starts recording as soon as it is configured
and will continue to do so until you start the unit via the button. No alarms are processed
while pre-start readings are being taken. Using pre-start readings is a good way to avoid
data loss if you forget to start the unit, as you can still access the data using LogTag
Analyzer.
Start delay
If you configure the TREL30-16 to start after a delay period, the logger will not
immediately record temperature readings after you have pressed
start a countdown timer instead, and record readings only after the timer has ended. The
value for the timer is set during configuration. If pre-start readings are enabled, these will
continue to be recorded during the delay period.
Date/Time start
The logger will start taking temperature readings at the date and time you enter during
configuration (local time). You cannot combine a date/time start with pre-start readings or
the start delay function.

Audible Alarm

The TREL30-16 is fitted with a buzzer. You can choose to activate the buzzer when an
alarm event has been triggered to provide extra feedback. This is enabled or disabled in the
Advanced Settings
Please note, that continual activation of the audible alarm will reduce the working
life of the battery. When an alarm event is triggered, the alarm should be cleared
as soon as possible.
TREL30-16 User Guide
®
Analyzer you can decide when the TREL30-16 starts
Starting the Logger on
Begin recording after a
when configuring the logger with LogTag
page 19).
Enable pre-start logging
delay.
®