Gui delines and manufacturers declarati on - Electromagnetic i mmunity
The Barkey p lasmatherm is intended for operation in one of the electro magnetic enviro n-
ments listed below. The customer or user of the Barkey plas matherm must ensure that it is
operated in one of these environments.
IEC 60601
Immunity test
test level
Conducted HF
3 V
e ff
interference ac-
150 kHz to
cording to IEC
80 MHz
61000-4-6
3 V/ m
Radiated HF in-
80 MHz to
terference accord-
2.5 GHz
ing to IEC 61000-
4-3
NOTE 1 At 80 M Hz and 800 MHz the higher frequency range applies.
NOTE 2 These guidelines may not apply in all cases. The propagation of electromagnetic
variables is affected by absorption and reflection of buildings, objects and people.
a
The field strength of static transmitters, such as base stations of radio telephones and
mobile land radios, amateur radios, AM and FM radio and telev ision transmissions cannot be
accurately theoretically determined in advance. You should consider carrying out a site survey
to determine the electro magnetic environ ment with regard to static transmitters. If the meas-
ured field strength at the location where the Barkey plasmatherm is being used exceeds the
above compliance levels, the Barkey p lasmatherm should be monitored to ensure that the
device is functioning as intended. If unusual performance characteristics are observed, add i-
tional measures such as changing the align ment or location of the Barkey p lasmatherm may be
necessary.
b
The field strength should be less than 3 V/ m in the frequency range 150 kHz to 80 M Hz.
Electromagnetic environment -
Compliance level
guideline
Portable and mobile rad io sets should
not be used within a distance fro m the
Barkey plas matherm, including cords,
that is less than the recommended
safety distance as calculated by the
relevant equation for the transmit
frequency.

Recommended safety distance

d
. 1
for
150
3 V
e ff
d
. 1
150 kHz to 80
for
80
Mhz
d
. 2
3 V/ m
80 MHz to 2.5
for
800
GHz
where P is the nominal power of the
transmitter in Watts (W) as stated by
the transmitter manufacturer, and d is
the recommended safety distance in
metres (m).
The field strength of stationary
transmitters should always be less
than the compliance level
frequencies in accordance with an on-
site investigation
Interference is possible near devices
that display the following symbol.
Safety advice
17
P
kHz
to
80
MHz
17
P
MHz
to
800
MHz
33
P
MHz
to
2.5
GHz
b
at all
a
.