Barkey plasmatherm Instructions For Use Manual page 62

Table of Contents

Advertisement

Error
Error description
number
10, 11
Display fault
20 to 23 Internal device error
30, 31
Checksum of the internal
device data is incorrect
32
Checksum of the program
memory is incorrect
33
Checksum of the data
memory is incorrect
40
Clock is incorrect
41
Error setting clock time
42
Error setting date
43
Error reading clock time
44
Error reading date
45
Clock not running
60
The watchdog timer has
detected an error.
70 to 73 Error with temperature
measurement / temperature
sensors
80 to 81 Overtemperature (soft-
ware) during temperature
measurement
90
Pump and heater supply
voltage not available.
91
Faulty program function
92
Overtemperature (elec-
tronic) during temperature
measurement
97
Faulty undulation function.
98
No power supply for pad-
dle motor.
100
Fill level in tank is critical;
heater has been switched
off
110
Communication / interface
fault
Comments
There may not be any text visible in the display.
The display may not be correct.
Replace battery
(by authorised specialists only).
Then set time and date.
Possible cause: sensor open circuit / short-
circuit
Restart system / allow device to cool down
Possible cause: overtemperature
Restart system / allow device to cool down
Restart system / remove paddle blockages
Restart system / top up the tank
Error messages