Safety advice
3.2

Safety advice on handling the device

Before using the device, carefully read and familiarise yourself
with these instructions and the user documentation for the op-
WARNING
tional accessories.
All electrical installations must comply with the relevant appli-
cable regulations and standards in addition to the specifications
stated by the manufacturer.
Only power supply connections supplied by Barkey GmbH &
Co. KG which are designed for the device's rated voltage may
be used.
The mains plug must be removed from the mains socket to
ensure safe isolation of the device from the power supply.
The device contains no parts which can be repaired by the user.
Do not attempt to repair the device yourself. You should con-
tact the manufacturer or your medical service technician who
can request information about repairs from the manufacturer if
necessary. Repairs and modifications to the device may only
be carried out by Barkey GmbH & Co. KG or authorised per-
sonnel.
The heating cushions of the device must not be allowed to
come into contact with sharp-edged objects.
The heating chamber and the heating cushions must be cleaned
and disinfected at least once per week! The filter paper must be
replaced after each cleaning.
An annual safety inspection must be carried out by qualified
service technicians or employees of Barkey GmbH & Co. KG.
The water must be changed once a year! You should always
add two micropur tablets when refilling.
CAUTION
Do not tilt the device when it is switched on!
The battery (lithium battery CR 1225, 3 V) must be replaced
every three years by qualified service personnel or employees
of Barkey GmbH & Co. KG.
Repairs and modifications to the device may only be carried
out by qualified service technicians or by employees of Barkey
GmbH & Co. KG.
The device's rating plate is on the left-hand side of the housing.