7.5 Test measurements

All the test measurements which can be displayed by pressing the TEST button are also
available to the host computer via the RS-232 interface. The host computer should issue
a request for a test measurement, and then the CPU will send the current value of the test
measurement to the RS-232 output. The format of the test measurement message is:
"T DDD:HH:MM IIII TEST MEASUREMENT<CRLF>"
For example, the format of the DC Power Supply output in millivolts would be:
"T 194:11:29 0000 DCPS= 2500 MV<CRLF>"
To request a test measurement, the host must issue a command of the form:
For a summary of all test functions issue the command "T LIST". The table below lists
the commands and the corresponding test measurements which will be returned.
TEST MEASUREMENT REQUEST COMMANDS
Command
"?<CRLF>"
"T LIST<CRLF>"
"T CO<CRLF>"
"T COMEAS<CRLF>"
"T COREF<CRLF>"
"T MRRATIO<CRLF>"
"T SPRESS<CRLF>"
"T SFLOW<CRLF>"
"T STEMP<CRLF>"
"T BNCHTEMP<CRLF>"
"T WHEELTEMP<CRLF>"
"T BOXTEMP<CRLF>"
"T DCPS<CRLF>"
"T COSLOPE<CRLF>"
"T COFFSET<CRLF>"
"T CLKTIME<CRLF>"
P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 62
"T MEASUREMENT<CRLF>"
TABLE 7.7
Test measurement
RS-232 HELP screen
Summary of all TEST's
Current CO reading
Current CO MEAS mV
Current CO REF mV
Current MR RATIO
Sample pressure
Sample flow rate
Sample temperature
Optical Bench temperature
Filter Wheel temperature
Internal box temperature
DC power supply output
Slope value
Offset value
Current time-of-day