Specifications - Teledyne 300 Instruction Manual

Carbon monoxide analyzer
Hide thumbs Also See for 300:
Table of Contents

Advertisement

1.4 Specifications

Ranges
Zero Noise
Span Noise
Lower Detectable Limit
Zero Drift (24 hours)
Zero Drift (7 days)
Span Drift (7 days)
Linearity
Precision
Lag Time
Rise/Fall Time (95%)
Sample Flow Rate
Temperature Range
Humidity Range
Temp Coefficient
Voltage Coefficient
Dimensions HxWxD
Weight
Power
Power, CE
Environmental Conditions
Recorder Outputs
Status
P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 4
User selectable to any full-scale range from 1 ppm
110V~/60Hz, 220V~/50Hz., 240V~/50Hz. 250 Watts
Installation Category (Overvoltage Category) II
Pollution Degree 2
± 100 mV, ± 1 V, ± 5 V, ±10 V (Bi-Polar)
12 status outputs from opto-isolators
* at constant temperature and voltage
to 1,000 ppm
< 0.025 ppm (rms)
< 0.5% of reading (rms)
< 0.050 ppm
* <0.1 ppm
* <0.2 ppm
* 1% of reading
1% FS
0.5% of reading
10 sec
<60 sec
800cc/min. ± 10%
5-40°C
0-95% RH, non-condensing
< 0.05 % per °C
< 0.05 % per V
7"x 17"x 25"
(178mm x 432mm x 660mm)
50 lbs (22.7 kg)
230V~/50Hz, 2.5 A

Advertisement

Table of Contents
loading

Table of Contents