To Compensate The Analog Probes; Figure 13 Example Pulses - Agilent Technologies 6000 Series Service Manual

Hide thumbs Also See for 6000 Series:
Table of Contents

Advertisement

4
Troubleshooting

To compensate the analog probes

78
If you see the waveform, but the square wave is not shaped
correctly as shown above, perform the procedure
compensate the analog probes"
If you do not see the waveform, ensure your power source is
adequate, the oscilloscope is properly powered-on, and the
probe is connected securely to the front-panel analog channel
input BNC and to the Probe Comp terminal.
6 If you still do not see the waveform, use the troubleshooting
flowchart in this chapter to isolate the problem.
You should compensate your analog probes to match their
characteristics to the oscilloscope's channels. A poorly
compensated probe can introduce measurement errors.
1 Perform the procedure
operation"
on page 77
2 Use a nonmetallic tool to adjust the trimmer capacitor on the
probe for the flattest pulse possible. The trimmer capacitor is
located on the probe BNC connector.
Perfectly compensated
Over compensated
Under compensated
Figure 13
Example pulses
3 Connect probes to all other analog channels (channel 2 of a
2-channel oscilloscope, or channels 2, 3, and 4 of a 4-channel
on page 78.
"To verify basic oscilloscope
comp.cdr
6000 Series Oscilloscopes Service Guide
"To

Advertisement

Table of Contents
loading

Table of Contents