To verify bandwidth
6000 Series Oscilloscopes Service Guide
This test checks the bandwidth of the oscilloscope. In this test
you will use a signal generator and a power meter.
1 GHz Models
Test limits at 2 mV/div to 5 V/div
• All channels (±3 dB)
dc to 1 GHz
•
500 MHz Models
Test limits at 2 mV/div to 5 V/div
• All channels (±3 dB)
dc to 500 MHz
•
300 MHz Models
Test limits at 2 mV/div to 5 V/div
• All channels (±3 dB)
dc to 300 MHz
•
100 MHz Models
Test limits at 1 mV/div to 5 V/div
• All channels (±3 dB)
dc to 100 MHz
•
Testing Performance
2
41