Agilent Technologies 6000 Series Service Manual page 38

Hide thumbs Also See for 6000 Series:
Table of Contents

Advertisement

2
Testing Performance
38
Table 7
Settings Used to Verify Voltage Measurement Accuracy
Volts/Div Setting
Oscilloscope
Calibrator Setting
5 V/Div
35 V
2 V/Div
14 V
1 V/Div
7 V
500 mV/Div
3.5 V
200 mV/Div
1.4 V
100 mV/Div
700 mV
50 mV/Div
350 mV
20 mV/Div
140 mV
10 mV/Div
70 mV
5 mV/Div
35 mV
1 (for 1 GHz, 500 MHz, and 300 MHz models)
2 mV/Div
14 mV
2 mV/Div
(for 100 MHz models)
14 mV
2
1 mV/Div
7 mV
1
Full scale is defined as 32 mV on the 2 mV/div range for 300 MHz, 500 MHz, and 1 GHz models.
2
Full scale is defined as 16 mV on the 1 mV/div range for 100 MHz models.
Full scale on all other ranges is defined as 8 divisions times the V/div setting.
c Press the Acquire key. Then press the Averaging softkey and
set #Avgs to 64.
Wait a few seconds for the measurement to settle.
2 Press the Cursors key, set the Mode softkey to Normal, then
press the X Y softkey and select Y. Press the Y1 softkey, then
use the Entry knob (labeled
the Y1 cursor on the baseline of the signal.
Test Limits
34.04 V
to
13.616 V
to
6.808 V
to
3.404 V
to
1.3616 V
to
680.8 mV
to
340.4 mV
to
136.16 mV
to
68.08 mV
to
34.04 mV
to
13.232 mV
to
13.616 mV
to
6.616 mV
to
on the front panel) to set
6000 Series Oscilloscopes Service Guide
35.96 V
14.384 V
7.192 V
3.596 V
1.4384 V
719.2 mV
359.6 mV
143.84 mV
71.92 mV
35.96 mV
14.768 mV
14.384 mV
7.384 mV