6000 Series Oscilloscopes Service Guide
5 To reinstall:
a Place the protective lens into the pocket in the display
mount and ensure that the inside of the lens is clean.
b Place the gasket into the slot around the lens making sure
it is fully seated in the slot all the way around.
c Clean the LCD window and insert the LCD under the
guides on the right side of the display mount.
d Push the left side of the LCD down until it clicks under the
latches so that the latches are fully over the face of the
sheet metal housing.
Figure 33
Latch over face of sheet metal housing
e Reinstall the backlight inverter (see
Replacing 6000A Assemblies
Sheet metal face
Latch
step 9
on
page
5
109).
113