Replacing 6000L Assemblies

To remove the front panel assembly

134
1 Perform the following procedures:
"To remove the bottom cover"
2 Insert a flat-blade screwdriver under the center of the
intensity knob and gently twist the screwdriver as you pull
the knob off.
Using a twisting motion on the screwdriver rather than
prying prevents marking or damaging the front panel.
Intensity
Knob
Figure 49
Removing the intensity knob
3 Disconnect the front panel ribbon cable from the system
board by pressing the release tabs and then disconnect it
from the front panel PC board.
4 Use a flat-blade screwdriver to release the center retainer tab
and push it slightly out of the slot.
5 Use a flat-blade screwdriver to release the 2 side retainer
tabs and remove them from their slots.
on page 133