To remove the LCD, gasket, and protective lens from the display
mount
Display Mount
6000 Series Oscilloscopes Service Guide
1 Perform the following procedures:
"To remove the cabinet"
"To remove the front panel assembly"
"To remove the display assembly"
"To remove the backlight inverter board and e-field
shield"
on page 107
2 Use a long-nose pliers or flat-blade screwdriver to push and
release the two LCD latches on the left side of the display
mount.
Release display mount latches
Figure 31
Replacing 6000A Assemblies
on page 94
on page 97
on page 103
LCD
LCD Latches
5