To remove the front panel assembly

6000 Series Oscilloscopes Service Guide
1 Perform the following procedures:
"To remove the cabinet"
2 Insert a flat-blade screwdriver under the center of the
intensity knob and gently twist it as you pull the knob off.
Using a twisting motion rather than prying prevents marking
or damaging the front panel.
3 Remove the T6 screws securing the BNC assembly to the
deck.
This step helps prevent the BNC connectors from binding
when removing and reinstalling the front panel.
Intensity
Knob
Figure 20
Removing the intensity knob and T6 screws
4 Disconnect the keyboard ribbon cable from the keyboard.
5 Use a flat-blade screwdriver to the release retainer tabs and
then push the panel forward.
Ensure that the retainer tab on the display side moves past
the rear edge of the display mount.
Replacing 6000A Assemblies
on page 94
T6 Screws (3 or 4)
5