Modbus Map - E+E Elektronik EE360 Operation Manual

Moisture in oil transmitter
Hide thumbs Also See for EE360:
Table of Contents

Advertisement

Menu
Modbus
settings*
Baudrate
Parity
Stop bits
Address
* Menu only available with connected Modbus RTU module during EE360 start-up.
Data transmission
Baud rate
Data bits
Parity
Stop bits
Slave address
The recommended setting for multiple devices in a Modbus RTU network is 9600 8 Even 1.
4.4

Modbus Map

Register
[DEC]
address [HEX]
Read registers: function code 0x03 / 0x04
31021
31003
31005
31009
31135
31141
Write registers: function code 0x06 for 16-bit and 0x10 (decimal: 16) for 32-bit
0001
26
value
None
Odd
Even
value
value
Factory settings
9600
300, 600, 1200, 2400, 4800, 9600, 19200, 38400, 57600, 76800
8
8
EVEN
None, odd, even
1
1 or 2
231
1...247
Protocol
Measured value
3FC
Relative humidity
3EA
Temperature
3EC
Temperature
3F0
Temperature
46E
Water activity
474
Water content
0
Slave-ID
Selectable values
Unit
Type
%
32-bit float
°C
32-bit float
°F
32-bit float
K
32-bit float
aw
32-bit float
x
32-bit float
/
16-bit integer
Operating instructions for EE360 Moisture in Oil Transmitter