VistOR SP 120 VAC
Switch, toggle
(Main power/fan)
20A 125 VAC
10A 250 VAC
Thermal fuse
(set at 152 C)
VistOR SP 230 VAC
Fuse
1.0 A 250 V
Slow Blow
Switch, toggle
(Main power/fan)
20A 125 VAC
10A 250 VAC
FIGURE 46: Electric Schematic for VistOR SP, 115/120 VAC & 230 VAC
5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410

VistOR SP Electric Schematic

115/120 VAC 50/60 Hz
Fuse
2.0 A 250 V
Slow BLow
20 VAC
230/240 VAC 50/60 Hz
20 VAC
Thermal fuse
(set at 152 C)
A member of the Medical Illumination International Group
Neutral
Fuse
2.0 A 250 V
Slow BLow
Toroidal Power
Transformer
Rated @ 20 VAC 170 VA
Lamp (Front)
150 W 20 V
Halogen type DDL
20 VAC
Fan / Motor
24 VAC, 50/60 Hz 17W
Neutral
Fuse
1.0 A 250 V
Slow Blow
Toroidal Power
Transformer
Rated @ 20 VAC 170 VA
Lamp (Front)
150 W 20 V
Halogen type DDL
20 VAC
Fan / Motor
24 VAC, 50/60 Hz 17W