Lamp Replacement VistOR PR and VistOR PRO

The VistOR lights operate at high temperatures. Allow the unit or units to cool at
least 30 minutes before performing any maintenance. Failure to do so could result in
personal injury.
Installation and repair of this equipment should be performed by qualified persons
only. Nuvo does not warranty any damage or injuries occurring as a result of improper
installation.
Unplug the power cord (floor models) before lamps are replaced. Failure to do so
could result in personal injury and/or property damage.
• For the VistOR PR and VistOR PRO pull off the rear housing cover from the light head as
shown in figure 36.
• Remove the lamp holder assembly from the light head by pulling it by the knob.
• Grasp the bulb by its base and carefully remove it from the lamp holder assembly.
• Replace the bulb with Nuvo, Inc. parts. For the VistOR PR and VistOR PRO use light bulb
P/N 0001256.
• Use the bulb's wrapping (plastic bubble sheet) to handle bulb and avoid skin contact.
Do not touch the lamp directly. Body oils may significantly lower the life expectancy
of the lamp and cause equipment damage. If contact is made, clean the lamp with
alcohol after installing it into the holder.
• Carefully reinstall the lamp holder assembly back into the light head. Care should be taken
not to bump the lamp against any internal components.
• Reinstall the rear housing cover.
5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410

Section 6: VistOR Maintenance

A member of the Medical Illumination International Group
Section 6
P
age 54 of 85