Figure 29: Ceiling Rod And Roll Pin Installation - Nuvo VISTOR Installation & Service Manual

Table of Contents

Advertisement

• Remove the ¼" by 2" roll pin taped down in the trolley assembly. Insert the pin into the
ceiling rod end that has one hole in both walls. The pin should extend the same amount on
both sides of the ceiling rod. Install the ceiling rod into the trolley and allow the roll pin to rest
in the indentation in the trolley. See figure 29.
Failure to install the roll pin correctly can cause serious injury and/or property
damage.
• Secure the ceiling rod by tightening the four ¼"-20 set screws in the trolley hub. Insert the
wire harness exiting the light head through the ceiling rod. The large Molex connector from
the wire harness must be located at the roll pin end. Insert the connector into the cord
bracket attached on the trolley's bottom until it locks in place.
Ceiling
Rod
Adjustment
Screws
Trolley
Assembly
Cord
Bracket
• Slide the trolley with the installed ceiling rod back onto the track. The cord-mounting bracket
in the trolley should be positioned so it is next to the mating Molex connector on the power
cord. Insure that the trolley rolls freely along the track. If adjustment is necessary loosen the
adjustment screws and position the side rollers against the base track extrusion as shown on
figure 27. Retighten the screws.
Ensure the power cord does not get twisted when attaching it to the cord bracket on
the next step. Do not proceed with the installation before correcting the problem.
5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410
Track Mount Installation Procedure

FIGURE 29: Ceiling Rod and Roll Pin Installation

A member of the Medical Illumination International Group
Side Rollers
Alignment Bracket
Roll Pin
Wire
Harness