• Remove sterilizable handle by pressing the button near the base of the handle and pulling
the handle off the handle post (see figure 43).
• Sterilize the handle utilizing steam sterilization of minimum 250° Fahrenheit for a minimum of
30 minutes in compliance with AAMI-SSSA-1988: Good Hospital Practices, Steam
Sterilization and Sterility Assurance, or equivalent method.
Sterilizable
Handle
P/N NV0001282

FIGURE 43: Handle Sterilization

Note: The VistOR EXL Exam/Diagnosis light is shown.