Nuvo VISTOR Installation & Service Manual page 32

Table of Contents

Advertisement

Improper fastening of the wall bracket can cause serious injury and/or property
damage. Make certain the installation is capable of supporting a load of at least 50 pounds
and an off center moment of 150 ft-lbs.
To maintain proper grounding reliability, the ground wire connections with the wall
bracket must be kept properly fastened at all times
Do not install the extension arm with the light head attached. Installation with the light
head attached can cause damage to the light. Refer to the procedure for installing the head
to the arm after the bracket and arm are installed.
• Attach the wall bracket to the desired location using two
(minimum length) lag screws or bolts with washers (Hardware not provided). See figure 12.
The bracket must be anchored to a structural backing with sufficient support. Failure
to use the correct backing could cause the arm/head to become unstable causing personal
injury and/or property damage.
• Remove the four screws holding the cord retainer plate in place. Carefully pull the plate away
from the wall bracket to expose the Molex connector. DO NOT use excessive force during
this procedure or in any way loosen or disconnect the green wires within the bracket.
Failure to comply with local electrical codes can cause a shock hazard.
• Bring the extension arm over the wall bracket and feed the Molex connector from the
transformer housing into the bracket while inserting the housing into the bracket. Secure the
housing to the bracket with the flat head cap screw provided.
• Bring the cord retainer plate up to the bottom of the wall bracket and plug the Molex connector
at the end of the power cord to the connector from the transformer. Route the connectors and
wire to the rear of the bracket (not shown). Refasten the retainer plate to the bracket with the
four screws provided.
NOTE: The light installation can also be wired to a power source permanently. The wall
bracket must first be mounted over an electrical junction box. Remove the hospital plug
from the power cord. The cord can then be inserted through the rear opening in the bracket
and wired directly to the power source. Wiring must be done in compliance with all
applicable electrical codes.
Failure to comply with local electrical codes can cause a shock hazard.
• See the instructions for "Installation of Light Head to Arm" on page 40.
5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410
Wall Mounted Light Installation
A member of the Medical Illumination International Group
5
/
" Diameter x 2 ½" long
16
Section 3
Page 32 of 85