2.5 Subscan Unit
Remarks for installation
• When reinstalling the sensor assy into the unit, while
pressing the detect roller of the V-sync sensor by finger as
shown in the right figure, slide the assy up. If the slide
roller is in upright position, it interferes with the activator
receptor plate, and cannot be slid in.
• When sliding the sensor assy up, align the assy to the left
end as close as possible.
• After confirming that the top edge of the sensor assy fits
into the sensor holder (leaf spring) of the main unit, tighten
the fixing screw.
72
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
Sensor Assy Holder
Sensor Assy
Fixing Screw