Konica Minolta REGIUS 190 Service Manual page 207

Hide thumbs Also See for REGIUS 190:
Table of Contents

Advertisement

A.1 SCB2 Dummy Program Screens
Item / Key
Casette
[Initialize]
receive
[Move to HP]
Unit
[Receive pos.]
[Transport pos.]
[Standby pos.]
Justify
[Initialize]
[Move to HP]
[Justify]
[Standby pos.]
180
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
Description
Initiates initialization of the Cassette receive Unit.
Moves the Cassette receive Unit to HP.
Moves the Cassette receive Unit to the receiving position where the cas-
sette is brought down when inserted.
Moves the Cassette receive Unit to the transport position where the cas-
sette is aligned to the level for transportation.
Moves the Cassette receive Unit to HP.
Initiates initialization of the justification assy.
Moves the justification assy to HP.
Initiates the justification operation (sandwiching the cassette) of the justifi-
cation assy.
Moves the justification assy to HP.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dd-941

Table of Contents