Replacement Of Sup1 (Dc5V / 24V Supply) - Konica Minolta REGIUS 190 Service Manual

Hide thumbs Also See for REGIUS 190:
Table of Contents

Advertisement

2.7 Electrical Unit
2.7.6

Replacement of SUP1 (DC5V / 24V supply)

1.
Detach the rear cover.
("2.2.1 Removal of the Rear Cover,
p.28")
2.
Remove a fixing screw "a", and loosen two fixing screws "b".
3.
Then remove the fan bracket.
Pay attention to the connector for DC supply cooling fan,
located on the back of the fan bracket, and do not apply
an excessive force.
4.
Unplug a relay connector of the DC supply cooling fan.
JJ60
5.
Unplug two connectors from SUP1 (DC5V / 24V supply).
JP2, JP3
6.
Remove 4 fixing screws, and pull out SUP1.
7.
Carry out reassembly in the reverse order to that described
above.
Remarks for installation
When it is replaced with a new SUP1, check the voltage at the
following point is DC 5V and DC 24V using a digital multi meter.
DC 5V: Between SUP1 JP3 / PIN 10 (5V) - PIN 7 (GND)
DC 24V: Between SUP1 JP3 / PIN 5 (24V) - PIN 2
(GND)
88
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
Fixing Screw "a"
Fixing Screw "b"
(remove)
(loosen)
SUP1 (DC5 / 24V supply)
Fan Blacket
Relay Connector