Konica Minolta REGIUS 190 Service Manual page 192

Hide thumbs Also See for REGIUS 190:
Table of Contents

Advertisement

Code
Medium Classifi-
cation
Signal process error
14010
Signal process error
14011
Detailed Clas-
sification
QRV Overflow
QRV Underflow
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
Countermeasures, etc.
The sensitivity has been mistakenly
corrected. Check the following.
1) Confirm that the ADB2's high volt-
age On/Off switch is On (check by
reading the solid density).
2) Confirm the absence of problems
in the exposure X-ray.
3) Replace the plate and confirm
operation.
4) Check the connectors in the vicin-
ity of the PMT and ADB2.
The sensitivity has been mistakenly
corrected. Check the following.
1) Confirm that the ADB2's high volt-
age On/Off switch is On (check by
reading the solid density).
2) Confirm the absence of problems
in the exposure X-ray.
3) Replace the plate and confirm
operation.
4) Check the connectors in the vicin-
ity of the PMT and ADB2.
4.4 Error Code
165

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dd-941

Table of Contents