A.3 Replacement Procedures of CF Card
8.
After the REGIUS Console and REGIUS 190 have successfully started up, start the service tool on the
REGIUS Console.
9.
On the "Network Setup" screen, restore the original settings (settings for the hospital) for the network of the
REGIUS 190 as it was installed.
10.
Restore the original settings (settings for the hospital) for the REGIUS Console as it was installed.
11.
Connect the REGIUS 190 and the REGIUS Console to the facility's network.
12.
Check that the all setting data except various correction data and config data of the REGIUS 190 has
restored the original status as it was installed.
13.
Start the service tool on the REGIUS Console, and restore the data onto the CF card of the REGIUS 190
using the "Board Change" screen.
Restore the data that was backed up in the step 1 or at the time of installation, servicing.
If there is no CF data available for restoration, check/readjust the REGIUS 190 according to the
"Procedures when no CF data is available"
14.
Restart the REGIUS 190.
Procedures when no CF data is available
When back up of the CF card is not successful, and there is no back up data available from the installation, it is
required to check/readjust as following after the replacement of CF card.
For the actual procedures of each work, refer to the "Installation/Service" manual.
• Checking the reading pixels, read start position.
• Adjusting the PLL setting, read start position.
• Unevenness correction.
• Sensitivity correction.
Important
The above procedures require tremendous steps to complete. Thus it is strongly recom-
mended to back up the CF card data at the time of installation.