4.4 Error Code

4.4 Error Code
4.4.1

General Procedures Required when Error Occurs

Below is the general procedure required when an error occurs.
Collect the log.
When an error is displayed on the LCD of REGIUS 190, collect the error log using REGIUS Console.
Reset
1.
Open / close the front cover.
2.
Remove the cassette if it is capable.
3.
Press "operation" button, and shut down the REGIUS 190.
4.
Turn OFF the circuit breaker located on the side of the main
unit.
5.
Restart the REGIUS 190.
Information to be attached when sending the error log
When sending the error log to the manufacturer, attach the related information listed below as detailed as possi-
ble.
Network Condition at the institute JM: External or internal?
Reset
Others, details
150
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
n: m
Any dedicated reader is connected?
IP address /subnet mask
REGIUS 190
JM
REGIUS Console
Reset implemented by opening / closing the door.
Needed to turn the breaker OFF / ON.