1.5 Description of Operation
11
22
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
•
The cassette ejection motor starts to operate and the cas-
sette is ejected.
•
If as a result of back plate detachment or the like, the cas-
sette passes the ejection shutter without a front plate, the
back plate sensor does not turn on and an error is gener-
ated.
•
If a subsequent cassette is inserted, drawing in of this cas-
sette will start when the ejection shutter close sensor turns
on.