A.3 Replacement Procedures of CF Card
CF card mounted on the SCB2 (System Control Board) contains the following data.
• Unevenness correction and shading correction data.
• Sensitivity correction data.
• Config data (PLL adjust value, start position/home position, etc.)
• Various setting data for networks.
Because the set of these data is unique to each REGIUS 190, it is necessary to back up the data before replace-
ment of the CF card, and to restore the data on the new CF card after the replacement.
Back up and restoration of these data can be implemented from the REGIUS Console that is connected to the
REGIUS 190 via network using the service tool.
Described below is the outline of replacement procedure of CF card. For the detailed operation of REGIUS Con-
sole (service tool), refer to "Installation/Service Manual".
<Caution> Replacement of the CF card shall be carried out when the error is attributed to the CF card. In
most of cases, network communication of the REGIUS 190 becomes disabled when an error
occurs on the CF. Therefore in addition to the back up of the CF data at the time of installa-
tion, which is a must, it is strongly recommended to back up the data whenever the setting
on the REGIUS 190 is changed at the time of maintenance work.

Replacement Procedures of CF Card

1.
If the network communication between the REGIUS 190 and REGIUS Console is valid, back up the data in
the CF card onto the REGIUS Console.
Start the service tool, and click "REGIUS190 --> REGIUS Console" of the "REGIUS 190 PCB" screen.
2.
Using one REGIUS Console (the one on which the data of the CF card has been backed up), back up the set
up data of this REGIUS Console onto the super disk. (hereinafter, implement the settings for REGIUS 190
and restore onto the CF card using this REGIUS Console)
3.
Restore the factory setting (default) onto the REGIUS Console.
On the "Network Setup" screen, restore the default settings. (IP address : 192.168.20.90, Sub-net mask :
255.255.255.0)
On the "JOB INFO" screen, put a check mark on [Job Manager(built-in)] of "Basic Setting", and "Manual
Register (afterward)" of "Barcode Setting".
On the "CCU INFO" screen, restore the default settings for [Host Name] of "TCP/IP", [Host Name (on Job
Manager)] and [IP Address]. ([Host Name] : R170-0001, [Host Name (on Job Manager)] : r170-0001, and
[IP Address] : 192.168.20.190)
On the "PostgreSQL" screen, restore the original settings (c_status : CS1-0001 only, r_status : r170-0001
only, relations : CS1-0001 and r170-0001, one relation only, sys_config : to "1", or on the "Service Tool"
screen, click "Adjust" --> [Job Manager] --> [Initialize] to initialize the settings.
4.
Disconnect the REGIUS Console and REGIUS 190 from the facility's network, and connect them in 1 to 1
configuration.
5.
Turn off the REGIUS 190 by pressing the operation button, and turn off the power breaker.
6.
Replace the CF card.
("2.7.2 Replacement of CF Card,
7.
Turn on the power breakers and start the REGIUS Console and REGIUS 190.
p.83")
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
A.3 Replacement Procedures of CF Card
187