Remote computer

To transmit data to a remote computer equipped with a modem:
1. From the SEND screen, select REMOTE COMPUTER.
2. At the PHONE NUMBER screen, enter the phone number for the remote
computer modem. For details, refer to Figure 5-4 and the instructions
immediately after it.
3. After entering the phone number, press SEND.
Some or all of these messages may be displayed:
DIAL REMOTE COMPUTER—the modem is dialing the number.
BUSY—the phone line is busy. The modem redials the number until the phone
line is no longer busy or until you press CANCEL.
SENDING DATA—data is being sent to the remote computer. A bar graph
shows the status of the transmission.

Figure 5-6. SENDING DATA screen

TRANSMISSION DONE—transmission is complete. The modem disconnects
the phone line. Press
If the transmission is interrupted, refer to Transmission failure messages later in
this chapter.

Wait for call

If your modem and the 3900/3900P are connected and powered on, the
3900/3900P is capable of automatically answering a call from a remote computer
and transmitting all data in trend memory. Intervention at the oximeter is not
required. The following procedure is necessary only if your modem does not
support this capability.
To transmit data when a remote computer initiates the transmission:
1. Select WAIT FOR CALL.
The following messages are displayed:
WAITING FOR CALL—the modem is waiting for an incoming call. You may
not use the 3900P printer while the modem is active. To cancel the
transmission, press CANCEL.
SENDING DATA—your modem received the call and is sending data to the
remote computer. Refer to Figure 5-6.
SENDING DATA ...
52%
to return to the SEND screen.
5/TeleOximetry
SpO
2
CANCEL
+
+
5-9