SETTINGS

Select SETTINGS from the main menu to access the first of two settings screens.
These settings are retained in memory and are not cleared when you clear trend
data or power off the monitor (as long as the internal battery is charged).

Figure 2-7. Settings screen options

Press the
Press the
respectively, through the options.
Press the – or
you have selected.
Press the
to the MENU screen.

SAVE LIMITS

Use this option to save the high and low SpO
pulse beep and alarm volumes you have set so they are retained when you power
off. The cursor is automatically on the SAVE LIMITS option field when the settings
screen appears. The options are NO and YES. To save the alarm limits settings for
use the next time the unit is powered on, select YES.
NOTE:
SAVE LIMITS:

ALL MUTE:

BAUD RATE:

ANALOG:

MODE:

OUTPUT:

CONTRAST:
function key to switch from screen to screen.
or
function key to move the cursor (a rectangle) down or up,
+
function key to cycle through the settings available for the option
function key to record your selections on these screens and return
Whenever a low SpO
to memory, a screen appears during the power-up sequence to alert the
operator. The screen shows the saved limit for the low SpO
saved alarm volume setting.
If your patients are neonates, use this option to save the appropriate high and
low alarm limits between power cycles.
NO
– / +
YES
9600
0.0V
SpO
2
– / +
6 SEC
5
and pulse rate alarm limits and the
2
alarm limit below 80% (or OFF) is set and then saved
2
2/Setup and Operations
SpO
+
2
+
+
SpO
2
+
alarm and the
2
2-15