Signal Noise; Figure 2-3. Low Perfusion Waveform; Figure 2-4. Noisy Plethysmographic Waveform - GE TruTrak+ 3900 User Manual

Table of Contents

Advertisement

CHECK SENSOR SITE

Figure 2-3. Low perfusion waveform

Signal noise

The following conditions can cause noisy waveforms:
Poor sensor placement.
Motion at the sensor site.
Electrical interference.

Figure 2-4. Noisy plethysmographic waveform

If three easily identifiable passes of a "typical" waveform do not occur,
Make sure the sensor's detector is flush with the sensor site (for sensor
application, see the instructions for the sensor you are using).
Make sure the light source and detector are directly opposite each other.
Select a site where the distance between the light source and the detector is less.
Make sure the patient site is stable; minimize movement of the sensor site.
Massage the sensor site with a 70% isopropyl alcohol pad or rubefacient cream
(10-30% methyl salicylate and 2-10% menthol) for 20-30 seconds. Strong
vasodilator creams, such as nitroglycerin paste, are not recommended.
If possible, remove electrical noise sources such as electrosurgery or
electrical/electronic devices (e.g., fans). If these solutions are not possible,
operate the oximeter on battery power, or try plugging the oximeter into a
different electrical outlet.
If artificial nails or excessive fingernail polish are present, select another site
or remove the polish/artificial nails.
2/Setup and Operations
2-9

Advertisement

Table of Contents
loading

This manual is also suitable for:

Trutrak+ 3900p

Table of Contents