Canon CXDI-60G Service Manual page 49

Table of Contents

Advertisement

3.2 Network setting for Screwcap.ini
The CXDI software communicates with the imaging part through Screwcap.dll by using the
communication protocol for sending and receiving commands and responses.
In conjunction with the communication, Screwcap.ini retains the information of the connected
imaging unit such as network addresses.
It is necessary to edit Screwcap.ini to communicate with the imaging part.
Since the Screwcap.ini factory setting is the following default setting (see the figure below), it is
not necessary to edit the setting unless you changed the network protocol TCP/IP setting for the
LAN card that communicates with the imaging unit in the previous item, "3.1 TCP/IP setting for
the control computer network card".
Screwcap.ini is located in the following directory:
D:\ccr\screwcap.ini
[HostInfo]
ResponseTimeout=30000
DataTimeout=60000
IntervalTimeout=30000
[SensorInfo]
SensorNum=1
[Sensor1]
IpAddress=192.168.100.11
CommandPort=12121
DataPort=12122
[Sensor2]
IpAddress=192.168.100.12
CommandPort=12121
DataPort=12122
[Sensor3]
IpAddress=192.168.100.13
CommandPort=12121
DataPort=12122
[Sensor4]
IpAddress=192.168.100.14
CommandPort=12121
DataPort=12122
[Log]
LogLevel=2
1
*
Imaging Unit IP address:
2. Installation
[SensorInfo]
SensorNum=1 Confirm this
Sensor IP address.
This must be the same as the
IP address stored in the
imaging part.
This address must be the same as the IP address stored in
the next item, "3.3 Network setting stored in the sensor".