< I n s p e c t i o n
ITEM
Category
5
System connections
6
Calibration
7
Self-diagnosis
Image quality
8
9
Transmission test
10
Ping test
11
PC-related items
12
System exterior
13
Final checks
NO
Item
1
Startup
CXDI Software
2
startup
3
Status lamps
4
Shutdown
X-ray generator
1
settings
2
Filter
Calibration
3
processing
Self-diagnosis
1
pass/fail
Phantom
1
exposure
1
Dcap, 200 times
Ping, 100,000
1
times
Dirt, dust inside
1
PC
Event logs
2
Application logs
Watson logs
System logs
3
OPU3 logs
4
Touch panel
5
Date, time
CXDI software
6
version
CCR folder
7
backup
1
Dirt
Hookup with RIS,
1
patient data
terminals
Exposure
information,
2
patient data in
film server
X-ray generator,
3
RIS and other
equipment
CXDI-60G/60C Inspection Check Sheet
I t e m s >
Characteristics values
The system power must
come on when the
power of the control PC
is turned on.
The system must start
up with no errors.
The lamps must provide
the correct indication
that corresponds to the
sensor statuses.
The system power must
go off when the power
of the control PC is
turned off.
The conditions under
which calibration is
implemented must be
noted.
Filter type
"Calibration completed
successfully."
Tests 1 to 5 must be
passed.
There must be no
artifacts, shading, grid
stripes, etc.
Transmission must be
successful.
Communication must be
successful.
There must be no dirt or
dust.
There must be no errors
(excluding errors which
have been dealt with).
There must be no errors
(excluding errors which
have been dealt with).
There must be no
deviation or other
problems.
The date and time must
be correct.
-
-
There must be no dirt.
It must be possible to
transmit and receive the
data correctly.
There must be no errors
in the information or
data.
The equipment must be
the same as before the
inspection work.
Page 2 of 2
Test means
Record
Visual check
Visual check
Visual check
Visual check
Visual check
msec
Visual check
Visual check
TEST1
TEST2
Visual check
TEST3
TEST4
TEST5
Visual check
Dcap.exe
Pingtest.exe
Visual check
Visual check
Visual check
Visual check
Visual check
Visual check
Ver.
External media,
etc.
Visual check
Visual check
Visual check
Visual check
kV
mA
Need help?