Replacing with the optional sensor cable
Cable clamp plate is not supplied with the optional sensor cable (S70-60/SP780-60) and
therefore you need to attach the cable clamp to the sensor cable by following the steps below.
1) Mount the cable clamp (lower) on the cable clamp plate.
* Make sure that the cable clamp (lower) is in the correct orientation.
2) Mount the sensor cable (optional) on the cable clamp (lower)
3) Hold the copper foil part of the sensor cable with the sensor cable clamp H60. Match the
edge faces of the copper foil and the cable clamp plate. Fasten them by the two screws
(M3x6mm).
* Fit the position of the Terminal No.1 of the P1 connector (V-shaped mark) to the position as
illustrated below. Fix the sensor cable.
* Be aware not to pinch or sandwich the wires.
*Reference value for the tightening torque: 12kgf·cm (117.6N·cm)
Sensor cable clamp plate H60
P2 connector
P1 connector
Position of the terminal No.1.
4) Fasten the cable clamp (upper) to the cable clamp plate by the two screws (M3x10mm).
* Reference value for the tightening torque: 8kgf·cm (78.4N·cm)
4. Repair Guide
(▼ mark)
Screws (M3x10mm) x2
Screws (M3x6mm)
Sensor cable (imaging unit side)
Cable clamp (Lower)
Cable clamp plate
6
S70-60/SP780-60
Cable clamp (Upper)